Loss of Atrial Signal Detection with Increasing Sensitivity Settings: A Software Problem and its Practical Solution
Gespeichert in:
Personen und Körperschaften: | , , , , |
---|---|
Titel: |
Loss of Atrial Signal Detection with Increasing Sensitivity Settings: A Software Problem and its Practical Solution |
In: | Pacing and Clinical Electrophysiology, 14, 1991, 4, S. 522-528 |
veröffentlicht: |
Wiley
|
Umfang: | 522-528 |
ISSN: |
1540-8159 0147-8389 |
DOI: | 10.1111/j.1540-8159.1991.tb02824.x |
Zusammenfassung: | <jats:p>A logic error in the software of Cosmos II results in apparent malsensing of atrial signals at high rather than low sensitivity settings. The paradoxical function was found in 54 out of 87 units tested (= 62%) if the atrial sensitivity was increased and the ventricular refractory period was lengthened beyond critical limits that could be defined individually. The problem inherent in both unipolar (66%) and bipolar systems (54%) is attributed to a summation of a physiological signal and internal noise, which has to be detected at the atrial level within a specific timing window to disable atrial sensing for the next pacing cycle. The malfunction virtually disappears if the ventricular refractory period does not exceed 210 msec. It is recommended, therefore, that Cosmos II pacemakers be left at their nominal ventricular refractory value (200 msec).</jats:p> |